Online BOD analyser

Continuous BOD at the outlet, when the laboratory test takes three days

Biochemical oxygen demand (BOD) is the oxygen micro-organisms use to break down organic matter. The laboratory test incubates a sample for three days at 27 °C, so it cannot be continuous. Online, BOD is derived from the same UV–visible spectrum as COD, using a BOD:COD correlation set at commissioning against your own laboratory results for that outlet.

  • SYZE-OCEMS-8P
  • One of 8 parameters as standard
  • CPCB
  • HSPCB
  • PPCB

At a glance

Supplied as part of the SYZE-OCEMS-8P, with all eight parameters as standard.

Parameter
BOD, biochemical oxygen demand
Range
0–1000 mg/L, set to your consented limits
Method
Derived from the UV–visible spectrum with a BOD:COD correlation
Instrument
The multi-parameter optical probe of the SYZE-OCEMS-8P, which also gives the other optical parameters
Reagents
None
Output
RS485 Modbus to the SYZE data logger
Reported to
CPCB, HSPCB, PPCB and PHED

Keeping BOD readings right

A correlation for your outlet
Set at commissioning from your laboratory's BOD and COD results for the same samples.
Rechecked when the effluent changes
A new process, a new industrial load or a change in treatment alters the BOD:COD ratio; the correlation is then checked again.
Reported with the others
BOD goes to the boards with COD, TSS, pH and flow, as the boards and the NGT order require.
Clean optics
BOD depends on the same optical window as COD, so it is cleaned on the same schedule.

Questions

Are online BOD readings accepted by the boards?

BOD derived from the spectrum is how continuous BOD is reported in practice. The station is registered on the portals as part of the installation, and commissioning is complete only when live data is seen there.

How often is the correlation checked?

At commissioning, at the intervals in CPCB's guidelines, and whenever the effluent changes or a laboratory check shows drift.

See it working

Send us a few recent laboratory BOD and COD results and we will tell you how the correlation is set.