How online COD, BOD and TSS analysers work

What the analyser measures, how it compares with the laboratory, and what affects accuracy

An online analyser does not run the laboratory test at the outlet. It measures something physical that follows the laboratory value - how the effluent absorbs and scatters light, or the voltage of a glass electrode - and does so every few minutes, all day. Knowing how that works explains most of the questions plants and boards ask about online readings.

Laboratory and online, side by side

ParameterIn the laboratoryOnlineWhat affects accuracy
CODDigestion with dichromate, about two hoursUV–visible absorption of the effluentA fouled window; an effluent different from the one it was set against
BODIncubation for three days at 27 °CDerived from the spectrum with a BOD:COD correlationA change in the effluent's make-up, which changes the BOD:COD ratio
TSSFiltration, drying and weighingOptical absorption and scatterAir bubbles, fouling
pHGlass electrodeGlass electrode with temperature compensationAn ageing electrode, a missed buffer calibration
TOCOxidation and measurement of the carbon dioxideUV–visible absorptionAs for COD
TurbidityNephelometerLight scattered at 90°Air bubbles, fouling

Why the correlation matters

A spectrum, not a number
The optical probe records how strongly the effluent absorbs light across the ultraviolet and visible range. COD, BOD, TSS and TOC are calculated from that spectrum.
Set for your outlet
Every effluent is a little different, so the calculation is set at commissioning against laboratory results for samples from your own outlet.
Checked over time
A grab sample tested in the laboratory is compared with the online value for the same time; a lasting difference means the setting is checked again.
No reagents
Because light does the measuring, the optical parameters need no chemicals and produce no waste.

What keeps the readings right

Clean optics
Fouling of the measuring window is the most common cause of drift; it is cleaned on a schedule.
Steady flow, no bubbles
The probe sits where the effluent flows steadily and bubbles do not pass the window.
A calibrated electrode
pH is calibrated with standard buffers and the electrode replaced when it ages.
A record of every check
Each calibration and laboratory comparison is recorded, with its date and values.

Questions

Is an online reading the same as a laboratory result?

It follows the laboratory value closely when the analyser is set and kept correctly, but it is a continuous measurement, not the same test. The two are compared regularly.

Why might the board's laboratory result differ from our online reading?

A grab sample is one moment; the online value may be an average over minutes. A small difference is normal. A lasting one is a reason to clean the probe and recheck its setting against the laboratory.